Calibration Service
The world class calibration facilities of the Field Spectroscopy Facility are available to academic and research institutes in the UK and Europe for the calibration of remote sensing field spectroradiometers. These services include the system response calibration of instruments for spectral radiance and irradiance measurements and the calibration of reference panels for spectral diffuse reflectance measurements. The facility also offers a wavelength validation service for quantifying the spectroradiometer's wavelength scale errors.
- Tables 1 & 2 - Calibrated Standards & Darkroom Equipment
- Tables 3 & 4 - Calibration Services & Costs
Calibration Standards & Equipment
Table 1 lists the National Physical Laboratory (NPL) calibrated standards used by the facility to perform the calibration services. Addition items of equipment used in the laboratory are listed in table 2.
Table 1 Calibration Standards
| Spectral Standard |
Equipment |
Spectral Range |
NPL Certificate No. |
| Spectral radiance source |
Hoffman LS-65-8D spectral radiance source |
300 - 2500 nm |
E04110364_B4-06 |
| Spectral irradiance source |
Optronic Laboratories 1KWatt FEL lamp |
250 - 2500 nm |
E09010185_B1-09 |
| Spectral reflectance panel |
Labsphere SRT-99-050 Spectralon target |
380 - 2500 nm |
E05090223 |
Table 2 Additional Equipment
| Equipment |
Services |
| Macam L203 photometer |
Independent verification of illuminance & luminance levels |
| Macam SMU203 spot measurement accessory |
Independent verification of radiance source luminance levels |
| Optronic Laboratories OL 83A stabilised current source |
Stabilised power supply for FEL irradiance standard |
| Mercury-argon spectral line soucre |
Wavelength scale verification (365, 404.7, 435.8, 546.1, ..... 912 nm) |
| Labsphere erbium oxide doped reflectance panel |
Wavelength scale verification (488, 522.2, 653.9, 799.2 .. 1534.4 nm) |
| NPL calibrated McCrone filter |
Wavelength scale verification (481, 588.6, 748.4, 869.2 .. 2326.8 nm) |
| NSIT calibrated Mylar sheet |
Wavelength scale verification (1128.7, 1659.9, 1904.6, ... 2444.5 nm) |
| Analytical Spectral Devices filter wheel & fibre source |
ASD FieldSpec Pro linearity and noise tests |
Calibration Services & Costs
Tables 3 and 4 list the services available from the Field Spectroscopy Facility for system response calibration and wavelength verification respectively.
Table 3 System Response Calibration for Spectral Radiance & Irradiance Measurements
| System |
Fore optics |
Calibration Service |
Cost (inc. VAT) |
| ASD FieldSpec Handheld |
Fixed 25° FOV Lens |
Spectral radiance 325 - 1075 nm |
XXX |
| ASD FieldSpec VNIR |
Fibre optic probe with one FOV lens |
Spectral radiance 350 - 1050 nm |
XXX |
| ASD FieldSpec VNIR |
Fibre optic probe with two FOV lenses |
Spectral radiance 350 - 1050 nm |
XXX |
| ASD FieldSpec VNIR |
Fibre optic probe with remote cosine receptor |
Spectral irradiance 350 - 1050 nm |
XXX |
| ASD FieldSpec FR & 3 |
Fibre optic probe with one FOV lens |
Spectral radiance 350 - 2500 nm |
XXX |
| ASD FieldSpec FR & 3 |
Fibre optic probe with two FOV lenses |
Spectral radiance 350 - 2500 nm |
XXX |
| ASD FieldSpec FR & 3 |
Fibre optic probe with remote cosine receptor |
Spectral irradiance 350 - 2200 nm |
XXX |
| Spectra Vista GER 1500 |
Standard or wide FOV lenses |
Spectral radiance 350 - 1090 nm |
XXX |
| Spectra Vista GER 1500 |
Standard & wide FOV lenses |
Spectral radiance 350 - 1090 nm |
XXX |
| Spectra Vista GER 1500 |
Standard lens with cosine diffuser |
Spectral irradiance 350 - 1090 nm |
XXX |
| Spectra Vista GER 1500 |
Fibre optic probe with cosine diffuser |
Spectral irradiance 350 - 1090 nm |
XXX |
| Spectra Vista GER 3700 |
Standard & wide FOV lenses |
Spectral radiance 350 - 2500 nm |
XXX |
| Spectra Vista GER 3700 |
Fibre optic probe with 25° FOV |
Spectral radiance 350 - 2500 nm |
XXX |
Table 4 Verification of spectroradiometer wavelength scale calibration
| System |
Fore optics |
Wavelength References |
Cost (inc. VAT) |
| ASD FieldSpec Handheld |
Fixed 25° lens |
Spectral line lamps, McCrone filter |
XXX |
| ASD FieldSpec VNIR |
Fibre optic probe with FOV lenses or RCR |
Spectral line lamps, McCrone filter |
XXX |
| ASD FieldSpec FR & 3 |
Fibre optic probe with FOV lenses or RCR |
Spectral line lamps, McCrone & mylar filters, erbium doped panel |
XXX |
| Spectra Vista GER 1500 |
Standard & wide FOV lenses |
Spectral line lamps, McCrone filter |
XXX |
| Spectra Vista GER 1500 |
Standard lens with cosine diffuser |
Spectral line lamps, McCrone filter |
XXX |
| Spectra Vista GER 1500 |
Fibre optic probe with cosine diffuser |
Spectral line lamps, McCrone filter |
XXX |
| Spectra Vista GER 3700 |
Standard & wide FOV lenses |
Spectral line lamps, McCrone & mylar filters, erbium doped panel |
XXX |
| Spectra Vista GER 3700 |
Fibre optic probe with 25° FOV |
Spectral line lamps, McCrone & mylar filters, erbium doped panel |
XXX |